Determination of plasma metformin by HPLC

Cai-cai JIN,Wen QIU,Hai-sheng JIAO
DOI: https://doi.org/10.3969/j.issn.1000-2812.2006.04.012
2006-01-01
Abstract:Objective To develop a quick and accurate HPLC method for the determination of metformin in human plasma.Methods Samples of plasma were deproteinated with acetoni- trile.Separation was achieved(5μm,4.6 mm×150 mm).The mobile phase consisted of acetoni- trile(35%),water(42.5%),methanol(15%),and sodium dodecyl sulphate(SDS,1.5%) at a flow rate of 1.0 mL/min.Metformin was detected by DAD detector at 234 nm.Results Standard curve was linear over the range of 0.154 to 2.316μg/mL.The detection limit in the plasma was 0.111μg/mL(s/n>3),while the recovery was more than 97%.RSD of intra-batch and inter-batch were all less than 8.0%.Conclusion The method is simple,rapid,and accurate,and can be used to determine the concentration of metformin in human plasma.
What problem does this paper attempt to address?