High performance liquid chromatography determination of low concentration metformin hydrochloride in beagle dog's plasma

Xi-tian PI,Hong-ying LIU,Xiao XIAO,Cheng-lin PENG,Xiao-lin ZHENG
2009-01-01
Abstract:Our objective is to develop a method to quantify low-concentration metformin hydrochloride in beagle dog plasma using High Performance Liquid Chromatography(HPLC).HPLC is conducted on a Dikma Diamonsil C18(150 mm×4.6 mm,5μm) with acetonitrile-10.0 mmol/L sodium dihydrogen phosphate(7.5 mmol/L SDS,pH5.5)(32:68) as a mobile phase,column temperature 30℃,detection wavelength 235 nm,and flow rate 1.0 mL/min.The internal standard method is used.The results show that the identified characteristics of metformin hydrochloride and the internal standard are distinct.The regression equation is Y =0.686 8X +(-0.022 2) r=0.999 5 and the linear range is 0.05~4.0 μg/L.The recovery rates are within 99.78%~107.90%.The intra-precision RSD is less than 3.85%,and the inter-precision is less than 5.63%.The developed method is sensitive,accurate,rapid and simple,and can be adopted to determine low-concentration metformin hydrochloride in beagle dog plasma and in pharmacokinetic study.
What problem does this paper attempt to address?