Determination of Plasma-Drug Concentration of Metformin Hydrochloride by RP-HPLC

QIAO Hua,CHANG Wei,WANG Ting,LIANG Li,ZHANG Xiao-yun
DOI: https://doi.org/10.3969/j.issn.1000-2812.2005.03.003
2005-01-01
Abstract:Objective To establish methods for determining plasma-drug concentration of met-formin hydrochloride by RP-HPLC. Methods metformin was isolated from plasma impurity by RP-RPLC, and plasma-drug concentration was calculated by linear regression between peak height of metformin and its drug concentration. Results Chromatographic condition for determining plasma-drug concentration of metformin hydrochloride by RP-HPLC was established. Its column is a 150 mm x4.6 mm column packed by Dikma with Diamonsil C18(5 μm), its column temperature is room temperature, its mobile phase is CH3CN to 0.02 mol/L KH2PO4 (containing 18% SDS of 0.32 mL per 100 mL, and adjust pH value to 4.0 with 85% H3PO4) in the proportion of 71 to 29, its flow rate is 1 mL/min, and ultraviolet wavelength for detection is 230 nm. The standard curve regression equation is C= 0.049 8H-0.00002, r= 0.9999, n =3 and the linear range is 0.030~4.000 mg/L. The recovery rate is 98.2%~99.0%. The intra-day RSD is 0.7%~4.6% and the inter-day is 1.2%~5.8%. Conclusion This method is simple and rapid in sampling, accurate and reliable in measurement, and can be applied for determination of plasma-drug concentration of metformin hydrochloride.
What problem does this paper attempt to address?