Determination of Tribendimidine and Related Substances in the Enteric-Coated of Tribendimidine by HPLC

孙艳华,黄桂华,肖扬帆,王洪波,刘云
DOI: https://doi.org/10.3969/j.issn.1672-7738.2010.07.010
2010-01-01
Abstract:OBJECTIVE To establish an HPLC method for the determination of tribendimidine in the entric-coated of Tribendimidine and its related substances.METHODS Chromatograghic column was Agilent Zorbax Extend-C18 (4.6×150mm,5μm),the mobile phase was acetenitrile-water-trieth (75:25:0.35),and detection wavelength was 388nm,the flow rate was 0.8mL·min-1.RESULTS A good linearity was abtained from 0.0253~0.1516mg·mL-1(r=0.9999),the average recovery was 99.38%.CONCLUSION The developed HPLC was proved to meet the requirement of detecting the content of the tribendimidine and related substances of the enteric-coated of Tribendimidine.The method had very good accuracy and precision,It was suitable for the quality control of the enteric-coated of Tribendimidine.