Analysis of Misdiagnosis of Coronary Artery Stricture with Electron Beam CT:A Report of 100 Cases

ZHAO Hong-liang,ZHENG Min-wen,ZUO Jian,HUAN Yi,SUN Li-jun,YOU Zhi-jun,XU Jian,PENG Yong
DOI: https://doi.org/10.3969/j.issn.1002-1671.2008.01.010
2008-01-01
Abstract:Objective To study the reasons of misdiagnosis of electron beam angiography(EBA)in coronary artery stenosis,and to improve the ratio of achieving image and the accuracy of diagnosis.Methods Coronary arteriography(CAG) and EBA were performed one after another in 100 cases suspected with coronary arteriosclerotic heart disease.4 coronary arterial branches to be showed on EBA were compared with that of CAG respectively,the misjudgement of the location of coronary arterial stenosis and the error of degree in judging arterial stenosis over 30% were defined as misdiagnosis.Results Among 400 coronary arteries of 100 cases,92 coronary arteries were misdiagnosed by EBA,the total misdiagnostic rate was 23%,of that,the misdiagnostic rate in the trunk of left coronary artery,the right coronary artery,the anterior descending coronary and circumflex branch of coronary artery was respectively 1.5%,4.8%,5.5% and 11.2%.The artifacts caused by heartbeat and respiration,arrhythmia,tachycardia,the unsuitable delay time and the masking by calcified plaque were all attributed to misdiagnosis.Conclusion To understand the reasons of misdiagnosis of coronary arterial stenosis by EBA is expected to improve image success rate and diagnostic accuracy.
What problem does this paper attempt to address?