The Experiment Study of Coronary Artery Imaging by 64 Slice Multidetector Computed Tomography Angiography with Pulsating Cardiac Phantom

霍天龙,杜湘珂,李剑颍,沈云,廖静敏
DOI: https://doi.org/10.3969/j.issn.1001-9324.2007.11.023
2007-01-01
Abstract:Objective To evaluate the diagnostic accuracy and reliability for detection of coronary artery stenosis and restenosis in stent using 64 slice multidetector computed tomography angiography(MDCTA).Materials and Methods A pulsating cardiac phantom(GE China CT Lab) with three simulated coronary arteries was scanned on 64 slice MDCT(GE Light Speed VCT) at static state and 3 different sinus rhythms of 50、70、90 beats per minute(bpm).Two simulated arteries were 3 mm and 5 mm respectively in lumen diameter with stenoses of 25%、50%、75%.A stent with stenoses of 50%、75% was placed inside the another artery with 4 mm in lumen diameter.Images were analyzed on the GE AW4.2 workstation and the results were compared with those true dimensions of simulated arteries.Results(1) The average measured values of the 25%、50%、75% stenoses and luminal diameters for the 3 mm and 5 mm simulated arteries were(0.61±0.02) mm、(1.11±0.06) mm、(1.68±0.05) mm、(2.28±0.06) mm(P values of 0.001、0.001、0.000、0.000 with one sample t test) and(1.14±0.04) mm,(2.30±0.04)mm,(3.50±0.04) mm,(4.71±0.04) mm(P values of 0.010、0.003、0.001、0.001 with one sample t test) respectively;The average measured percentage values of the 25%、50%、75% stenoses were(26.8±0.6)%、(48.6±0.7)%、(73.8±0.9)%(P values of 0.009、0.145、0.075 with one sample t test) and(24.1±0.6)%、(48.9±0.8)%、(74.4±0.7)%(P values of 0.066、0.071、0.139 with one-sample t test).(2) The measured values of the 3 mm diameter simulated artery were more susceptible to high heart rate with artifact,especially when they were more than 70 bpm.(3) 64 slice VCT had a potential but limited capability to depict in stent restenosis due to the artifacts caused mainly by high density metal stent and motions at high heart rate;for the same reason the diameter of lumen at stent segment was less than that of normal segment,which was statistical significant.(4) Measurements revealed significantly better results on the gantry rotation time of 0.42 s compared to that of 0.35 s(P=0.036) at heart rate of 90 bpm.Conclusion 64 slices VCT is reliable in detecting coronary artery stenosis,it is limited to depict in stent restenosis,VCT is promising in the cases when vessel caliber is relatively large and heart rate is relatively low.Improving the spatial resolution through software kernels or hardware can increase the accuracy in measuring in stent restenosis.Making fully usage of matched scan parameters can also do much help to imaging quality.
What problem does this paper attempt to address?