Concentration and Optical Measurement of Aqueous Analytes in an Organic-Solvent Segmented Capillary under High-Electric-Field

HJ ZHENG,PK DASGUPTA
DOI: https://doi.org/10.1021/ac00094a023
IF: 7.4
1994-01-01
Analytical Chemistry
Abstract:Concentration by solvent extraction is examined in a fused silica capillary under high electric field. An immiscible solvent segment, containing a sufficient concentration of a supporting electrolyte, can be introduced in an aqueous electrolyte-filled capillary by electroosmotic suction. Following the introduction of the solvent segment, the capillary terminus is reintroduced into an aqueous electrolyte and the electric field is restored. The segment is carried along by the electroosmotic flow. Aside from any intrinsic conductivity of the saline immiscible solvent, a thin aqueous film around it permits the electrical circuit to remain continuous. If ions or ionizable complex moieties (extractable as ion pairs) are initially present in the aqueous electrolyte, they are concentrated by extraction in the organic solvent and/or accumulate in the aqueous phase ahead of the interface. For the analyte extracted into the organic segment, the charge type and nature of the extracted ion and the applied voltage polarity determine whether the accumulation occurs on the front or the rear edge of the solvent segment. The peak analyte concentration in the organic segment is typically an order of magnitude greater than the initial concentration in the aqueous phase. If a similar experiment is conducted with the introduction and transport of an organic solvent segment under gravity, no significant extraction signal is observed.
What problem does this paper attempt to address?