Abstract:In this work, chloride, chlorate and perchlorate are fast separated on PDMS microchip and detected via in-channel indirect amperometric detection mode. With PDMS/PDMS microchip treated by oxygen plasma, anions chloride (Cl−), chlorate (ClO3−), and perchlorate (ClO4−) are separated within 35s. Some parameters including buffer salt concentration, buffer pH, separation voltage and detection potential are investigated in detail. The separation conditions using 15mM (pH 6.12) of 2-(N-morpholino)ethanesulfonic acid (MES)+l-histidine (l-His) as running buffer, −2000V as separation voltage and 0.7V as detection potential are optimized. Under this condition, the detection limits of Cl−, ClO3−, and ClO4− are 1.9, 3.6, and 2.8μM, respectively.