Electrochemical Detector for Microchip Electrophoresis of Poly(dimethylsiloxane) with a Three-Dimensional Adjustor.

N Bao,JJ Xu,YH Dou,Y Cai,HY Chen,XH Xia
DOI: https://doi.org/10.1016/j.chroma.2004.04.036
2004-01-01
Abstract:This paper presents an electrochemical detector for poly(dimethylsiloxane) (PDMS) microchip CE with a three-dimensional adjustor which makes it possible to accurately align a separate working electrode that can be easily fabricated in laboratory to the uncertain PDMS microchannel outlet. The substantial influence of the electrode—PDMS microchannel distance was investigated. The optimal electrode-outlet distance was found to be 10 μm for the PDMS microchannel with the width of 50 μm due to its relatively slow electroosmotic flow. Adrenaline and catechol were well separated, with a linear response range from 20 μM to 1 mM, and a detection limit of 2 μM for catechol, using carbon disk electrode (diameter of 300 μm). Furthermore, arginine and histidine can be well separated and detected directly in the PDMS microchannel using a Cu disk electrode (diameter of 150 μm).
What problem does this paper attempt to address?