A three-dimensionally adjustable amperometric detector for microchip electrophoretic measurement of nitroaromatic pollutants

Xiao Yao,Joseph Wang,Luyan Zhang,Pengyuan Yang,Gang Chen
DOI: https://doi.org/10.1016/j.talanta.2006.01.003
IF: 6.1
2006-01-01
Talanta
Abstract:A microchip capillary electrophoresis (CE)–amperometric detection (AD) system has been fabricated by integrating a two-dimensionally adjustable CE microchip and an amperometric detection cell containing a one-dimensionally adjustable disc detection electrode in a Plexiglas holder. It facilitates the precise three-dimensional alignment between the channel outlet and the detection electrode without a complicated three-dimensional manipulator. The performance of this unique system was demonstrated by separating four nitroaromatic pollutants (nitrobenzene, 2,4-dinitrotoluene, 2,4,6-trinitrotoluene, and p-nitrobenzene). Factors influencing their separation and detection processes were examined and optimised. The four analytes have been well-separated within 120s in a 75cm long separation channel at a separation voltage of +2000V using an electrophoretic separation medium containing 15mM borax and 15mM sodium dodecyl sulfate (pH 9.2). Highly linear response is obtained for the four analytes over the range of 0–5ppm with the detection limits ranging from 12 to 52ppb. The present system demonstrated long-term stability and reproducibility with relative standard deviations of less than 5% for the peak current (n=9). The new approach for the microchannel–electrode alignment should find a wide range of applications in other microfluidic analysis systems.