Stress Analysis of Single-ended Fixed Mandibular Prosthesis

GONG Zhen-yu,LI Guo-hua,XU Rui-jie,LI Di-chen,ZHOU Bing
DOI: https://doi.org/10.13241/j.cnki.pmb.2011.12.055
2011-01-01
Abstract:Objective:To analyze the stress and strain on titanium framework,end of mandible and fixing screws in load when the defect area includes unilateral condyle.Methods:Firstly establish 3D finite element model of defective mandible and titanium frameworks,and then vertical load were exerted on them to simulate the mastication movement.Finally,stress and strain situation of the model were analyzed.Results:When the model was loaded,the stress concentrated at base of extending plate and neck of the screw upper near bone end.Neck of the screw lower near bone end and lingual part farther from bone end deformed most severely.Conclusion:In design of titanium framework fixed unilaterally to repair the mandible bone defects including a condyle,the extend plate should be widen and overcoat the inferior border of it,moreover,stronger fixation in inferior border of mandible is necessary.