Study On Insertion Thermal Gas Mass Flow-Meter

徐国强,钟骏意
DOI: https://doi.org/10.3969/j.issn.1006-883X.2007.09.004
2007-01-01
Abstract:The design and calibration projects of a new insertion thermal gas mass flow-meter are offered in this paper. The principle and methods of insertion thermal gas mass flow-meter are expounded,and a simple flux expression with extensive application sphere is obtained while dealing with the criterion relation. The consideration factors of precision and calibration in manufacture are discussed. In addition,the measurements are proposed to reduce the heat loss,based on the thermal balance analyses. The flow-meter is calibrated with direct measurement method. Experimental results show that the flow-meter response is greatly improved using graphite as filler which has high thermal conduction coefficient density ratio. The flow-meter has perfect performance and can accurately measure gas mass flux.
What problem does this paper attempt to address?