Comparison of Imaging Quality and Radiation Dose Between Flat Detector DSA and II/TV DSA in Fluoroscopic Application

WANG Zhi-kang,SUN Jian-zhong,ZHANG Wei-min
DOI: https://doi.org/10.3760/cma.j.issn.0254-5098.2006.04.030
2006-01-01
Abstract:ObjectiveTo evaluate and compare the difference in image quality and radiation dose between flat detector(FD) DSA(PHILIPS Allura Xper FD20, Holland) and II/TV DSA (PHILIPS Integris CV12, Holland) in fluoroscopic application. Methods Based on threshold contrast_detail detectability (TCCD) technique, phantom T0 10 was tested by both systems with the following entrance air kerma(EAK) levels at the X ray detector entrance plane: 5, 10, 20, 40, 60 and 80 nGy/frame in fluoroscopic application. Three observers scored each fluoroscopy image sequence using double blind method, calculated the threshold detection index(H_T), and drew the detection index figures. Results Both systems have the same H_T value when EAK level was 5 nGy/frame.As the EAK level was increased, the H_T value of FD DSA system increased broadly, but the II/TV system increased slowly. However the H_T value of the II/TV system decreased if the EAK level was increased too high. Radiation dose of the FD DSA was less than half of the II/TV DSA radiation dose during clinical application at the same H_T value. ConclusionImage quality of the FD DSA is better than II/TV DSA. The FD DSA will reduce the radiation dose to the patient and doctor greatly comparing with II/TV DSA at the same image quality.