Determination of Thiamine in Pharmaceutical Preparations by Hexacyanoferrate(Ⅲ) Chemiluminescence System

CHEN Long,ZHANG Yan-tu,QI Guang-cai,LI Xiao-xia,CHEN Xiao-li,ZHANG Zhu-jun
DOI: https://doi.org/10.3969/j.issn.1004-602x.2006.04.017
2006-01-01
Abstract:A novel chemiluminescence(CL) system for thiamine combined with flow injection analysis was proposed in this paper.It was based on the reaction between thiamine and hexacyanoferrate(III) in sodium hydroxide medium to give CL signal,which was enhanced by cetyltrimethyl ammonium bromide(CTAB).The experimental conditions for the CL reaction were optimized.Under the optimum conditions,the concentration of thiamine was proportional with the CL intensity in the range of 0.06~360mg/L with a RSD of 1.44%(0.6 mg/L,n=11).The detection limit was 0.045mg/L(3σ).The recovery was in the range of 96.0%-103%.The method has been successfully applied to the determination of thiamine in pharmaceutical preparations.