Determination of vitamin B12 in multivitamin tablets and fermentation medium by high-performance liquid chromatography with fluorescence detection

H B Li,F Chen,Y Jiang
DOI: https://doi.org/10.1016/S0021-9673(00)00724-X
IF: 4.601
2000-01-01
Journal of Chromatography A
Abstract:A novel method for the determination of vitamin B12 by high-performance liquid chromatography with fluorescence detection is reported. The method was simple and highly sensitive with good precision. Vitamin B12 was analyzed by HPLC on a μBondapak C18 column (300×3.9 mm, 10 μm) with methanol–water (30:70) as mobile phase and fluorescence detection at 305 nm (with excitation at 275 nm). The calibration graph was linear from 1.000 to 100.0 ng ml−1 for vitamin B12 with a correlation coefficient of 0.998 (n=6). The detection limit was 0.1 ng ml−1. The method was successfully applied to the determination of vitamin B12 in vitamin B12 tablets, multivitamin tablets and fermentation medium. The recovery was from 94 to 102% and the relative standard deviation was in the range of 1.8 to 4.1%.