Cryogenic Microcalorimeters for High Resolution Spectroscopy: Current Status and Future Prospects

D MCCAMMON,W CUI,M JUDA,P PLUCINSKY,J ZHANG,RL KELLEY,SS HOLT,GM MADEJSKI,SH MOSELEY,AE SZYMKOWIAK
DOI: https://doi.org/10.1016/0375-9474(91)90238-2
IF: 1.558
1991-01-01
Nuclear Physics A
Abstract:Small thermal calorimeters operated at temperatures below 1 K can detect single photon or particle events that deposit only a few eV in the device. Their energy resolution for soft X-rays is potentially better than 1 eV FWHM, maintaining the high efficiency and large angular acceptance of a Si(Li) solid state detector with a factor of about 100 improvement in resolution. The best performance obtained so far is 7.3 eV FWHM for 6 keV X-rays with a 1/4 mm x 1 mm device operated at 60 mK. Other unique properties of microcalorimetry include sensitivity to non-ionizing events and a high tolerance for the inclusion of many source materials within the detector.We discuss fundamental factors influencing resolution, limitations of the current approach, applicability considerations, and possible routes for future improvements.