DETERMINATION OF TRACE THALLIUM IN SOIL SAMPLES BY GFAAS WITH POLYURETHANE FOAM

Wang Long-shan,Gao Deng-feng,Mu Nai-cang,Wang Xiao-yan
DOI: https://doi.org/10.3969/j.issn.1001-6996.2008.02.012
2008-01-01
Abstract:A method for the determination of trace thallium in soil samples by graphite furnace atomic absorption spectrometry after preconcentration with polyurethane foam has been developed.Digested sample in hydrochloric acid,nitric acid,perchloric acid and Hydrofluoric acid,trace thallium is adsorbed by polyurethane foam from 10% nitric acid medium in the presence of Fe+3 ion and released in 2ml hydrogen peroxide.Tl is determined by GFAAS with ascorbate as a matrix modifier.The detection limit of the method is 0.2 ng/g for thallium.The method has been applied to the determination of trace Tl in soil standard reference samples and the results are in good agreement with certified values as well as precision of 5.05% ~8.74% RSD(n=12).
What problem does this paper attempt to address?