Characters and Image Features of Micro Self-focusing Ultrasound Transducer Used to Treat Cervical Lesions

WANG Yi-ni,HUI Chun,SHOU Wen-de,ZHANG Yan-li,YU Li-li,GAO Shun-hua,XU Ai-lan,CHEN Ya-zhu
2009-01-01
Abstract:Objective: Design an annular self-ultrasound transducer with concave spherical surface used to treat cervical lesions, and analyze its focusing performance. Methods: Axial sound field distribution of annular self-ultrasound transducer with concave spherical surface used to treat cervical lesions is simulated by introducing a negative sound source. And exercise optimization design with the physical meaning of sound intensity distribution coefficient (SIDC). Then the focusing performance is verified by image feature analysis in beef liver experiment. Results: By the premise of mechanical robustness and their own design constrains, optimization design makeⅠtransducer (inner ring diameter; 6 mm; outer ring diameter: 12 mm) and Ⅱ transducer's (inner ring diameter: 3 mm; outer ring diameter: 7.5 mm) SIDC values reach maximum. The beef liver experiment results show that the axial length of lesion region near the focus was consistent with the simulation results; As Ⅱ transducer's SIDC value is bigger than Ⅰtransducer's, the damage region of Ⅱ transducer is bigger than it ofⅠtransducer and distributed more uniformly. Conclusions: Under the same condition, the transducer with bigger SIDC value has longer the axial length of lesion region near the focus, and distributed more uniformly.