Optimization of the focused acoustic field of self-focusing transducer by element control
Lei WU,Rui WANG,Yi-hui GAO,Xiao-ming GENG,Li-xin JIANG
DOI: https://doi.org/10.16300/j.cnki.1000-3630.2016.06.017
2016-01-01
Abstract:According to the characteristics of the focusing transducer using 6-element self-focusing mode in HY2900 focused ultrasound tumor treatment system, through sorting and quantity control of the transducer, the system is able to control the focused acoustic field to avoid special parts. The sound field distribution of the transducer in deaerated water is measured by Reyleigh’s integral calculation and the software HY2900. The order of 6-element transducer is defined, and the number of controlled elements and their order are determined. The sound field distributions on the focal plane and on the plane at 10 mm above the focal point are measured in four different conditions: transducer elements all driven, closing the 1, 6 element, closing the 1, 2, 6 element, closing 1, 3, 5 element. It is found that the acoustic distribution plane at 10 mm above the focal point directly reflects the number and the order of transducer elements causing the change of transducer focusing path. The optimization of the acoustic field distribution can be achieved by controlling the order and number of the focusing transducer elements.