The effects of pulse duration on efficacy of ventricular defibrillation

GU Yun-fei,JIN Jun,XUE Shu-feng,CHEN Jian-feng,HE Hui-juan,DUAN Ka-dan,ZHANG Shou-yan
DOI: https://doi.org/10.3969/j.issn.1006-5725.2013.02.036
2013-01-01
Abstract:Objective The aim of this study was to evaluate the effect of pulse duration on efficacy of ventricular defibrillation using automated external defibrillator(AED).Methods The pig's ventricular fibrillation(VF) was induced through alternating current(AC),then the pigs randomly received defibrillation with biphasic waveforms which pulse duration was 4 ms + 1 ms + 4 ms or 5 ms + 1 ms + 5 ms.Defibrillation thresholds(DFTs) and success rates were determined and analyzed by statistics.Results There are induced VF 80 episode and 25 achieve success.The two groups have no significant difference on total shock voltage,energy and electric quantity.But the defibrillation threshold of voltage,energy and electric quantity of 4 ms + 1 ms + 4 ms group were lower compare with 5 ms + 1 ms + 5 ms group(254.337 ± 93.191V Vs 310.818 ± 86.534V,P = 0.023;9.811 ± 5.167J Vs 11.340 ± 8.086 J,P =0.047;0.031 ± 0.011C Vs 0.037 ± 0.010C,P = 0.022).Conclusion The pulse duration is an important influence factor for defibrillation efficacy and we may decrease DFTs by shorten the pulse duration.