[Defibrillation Efficacy of a Newly Developed Automated External Defibrillator in a Swine Ventricular Tachycardia/fibrillation Model].

Jie Hui,Yun-Fei Gu,Xiao-Mei Wu,Zong-Cheng Zhu,Jian-Ping Song,Xiang-Jun Yang,Zhi-Hua Liu,Ting-Bo Jiang,Zu-Xiang Fagn,Wen-Ping Jiang
DOI: https://doi.org/10.3760/cma.j.issn.0253-3758.2009.09.018
2009-01-01
Abstract:OBJECTIVE:The aim of this study was to evaluate the recognition and defibrillation efficiency of a newly developed automated external defibrillator (AED).METHOD:Ventricular tachycardia (VT)/ventricular fibrillation (VF) was induced by alternating current (50 Hz) through an electrode placed on apex of right ventricle in 23 anesthetized swine and recorded, recognized and defibrillated by a newly developed AED.RESULTS:A total of 96 VF was induced and 145 defibrillations were recorded. We analyzed available 167 electrocardiosignal with a total length of 103,740 seconds. The accuracy, sensitivity and the specificity of the AED on VT/VF recognition are 99.5%, 98.2% and 99.6%, respectively. The success rate of defibrillation was 33.4% which increased in proportion to defibrillation energy. The defibrillation threshold of energy is 29.10-116.91 (78.75 +/- 35.64) J, the defibrillation threshold of electric quantity is (0.11 +/- 0.04) C and the defibrillation threshold of voltage is (1216.67 +/- 260.87) V.CONCLUSIONS:This newly developed AED has high sensitivity and the specificity on recognizing VT/VF. The lower success rate of defibrillation of this AED is associated with the low defibrillation energy during defibrillation which needs to be improved on further development.