Does the successful ventricular defibrillation decide by energy or charge?

Xiaomei Wu,Zuxiang Fang,Cuiwei Yang,Hailang Song
2007-01-01
Abstract:Objective: This paper investigates the main electrical parameters which influent the results of the ventricular defibrillation. The research is the foundation of low energy defibrillation. Methods: According to the cell electrophysiology, the change of charge is the main reason of the change in transmembrane potential. In this paper, the effection of the energy especially charge to defibrillation is studied through 7 animal trails by an appropriative defibrillator designed by ourselves. Results: The defribrillation energy is the main parameter which determines the dysfunctions to the patient and the defibrillation charge is the main parameter which determines whether ventricular defibrillation is successful. According to the defibrillation energy threshold curve and charge threshold curve gained from the animal trials, the shock pulse can be optimized in order to decreas the body dysfunctions as well as defibrillating successfully. The decisive infection of charge to defibrillation is brought forward in this paper for the first time.
What problem does this paper attempt to address?