The Medical Linac Photon Angular Distribution Study By Monte Carlo Method

XU Yun,WU Zhang-wen,HOU Qing,GOU Cheng-jun
DOI: https://doi.org/10.3969/j.issn.1005-202X.2013.02.001
2013-01-01
Abstract:Objective: In radiation therapy,the photon angular distribution influence the dose distribution and so on.Research the photon angular distribution of the medical linac for improving the accuracy of radiotherapy.Methods: BEAMnrc is built on the EGSnrc Code System and used to model the radiotherapy source which was developed 3-D treatment planning for radiotherapy.Use the BEAMnrc code to simulate the transport of electron and photon in the treatment head of the medical linac,where a phase space file is generated in SSD=100 cm.Analyze the phase space which contains the information of each one particle and obtain the photon angular distribution of the medical linac by using the BEAMdp code.Results: About the 6 MV linac,the photon angular distribution is the same when the analysis regions are the same in the different fields.For different off-axis area,the photon angular distribution and the cone angle of the area are the same basically.So the photon angular distribution can be estimated by the cone beam.Conclusions: The photon angular distribution and the analysis regions are relevant;The influence of the JAWS is little.In order to improve the life quality of the patients,we should consider the photon angular distribution carefully in radiotherapy dose calculation.
What problem does this paper attempt to address?