Preliminary study of real-time three-dimensional echocardiographic planimetry of the mitral valve area before and after percutaneous balloon mitral valvuloplasty

ZHANG Yan,TANG Hong,HUANG He,SONG Hai-bo,张嬿,唐红,黄鹤,宋海波
DOI: https://doi.org/10.3760/j.issn:1004-4477.2006.02.004
2006-01-01
Abstract:Objective To assess the feasibility and accuracy of real-time three-dimensional echocardiographic(RT3DE) planimetry of the mitral valve area(MVA) before and after percutaneous balloon mitral valvuloplasty(PBMV).Methods MVA was estimated in 18 patients before and after PBMV,using two-dimensional echocardiography(2DE),Doppler pressure half-time(PHT) and volumetric RT3DE.Results ①There were significant difference between MVA measured before and after PBMV(P<(0.001)),but MVA measured by RT3DE had no significant difference and correlated well with MVA measured by 2DE and PHT before(r_1=(0.845),r_2=(0.596)) and after PBMV(r_1=(0.753),r_2=(0.848)).Meanwhile MVA increases(ΔMVA)measured by RT3DE had no significant difference with that measured by 2DE and PHT(P>(0.05)).②MVA measured by RT3DE was slightly less than that measured by 2DE.③There were no significant difference among MVA measured by RT3DE in three zoom images and full volume image(P>(0.05)).Conclusions RT3DE planimetry of MVA before and after PBMV is feasible and accurate.It provides a new quantitative way for MVA following up after PBMV.
What problem does this paper attempt to address?