Absolute measurement of activity of 182Ta using the 4πβ+4πγ counting meathod

ZHANG Xiao-lin,LI Yuan-jing,LI Qi,FAN Jin-long,LIU Jie,CHENG Yong-fu
DOI: https://doi.org/10.3969/j.issn.0258-0934.2007.03.023
2007-01-01
Abstract:182Ta emits two groups γ-rays in the low and high energy range.And it has a moderate long half-life,so that it is one of the suitable standard sources to calibrate the efficiency curve of HPGe detectors.The specific activity of 182Ta solutionwas measured using the 4πβ+4πγ counting meathod.Sources of 182Ta were produced by thermal neutron in the reactor.The total efficiency of 4πβ+4πγ counting deviceto 182Ta reaches above 99%,and the relativetly combined standard uncertainty of the specific activity is 0.2%.
What problem does this paper attempt to address?