Low-dose MSCT:Application to Three-dimension Imaging of Central Airways

Ding Juan,Xiao Xiangsheng,Li Huimin
DOI: https://doi.org/10.3969/j.issn.1006-5741.2006.01.006
2006-01-01
Abstract:Purpose:Comparing of low and conventional dose MSCT of difference of reconstructive imaging quality of the central airway,then assess the utility of low-dose MSCT for three-dimensional imaging of the central airway.Materials and Methods:MSCT was performed in32patients using two tube currents(50and195mAs).4groups SSD and VR images were reconstructed with standard and high resolution.They were assessed by2radiologists for image quality and results were analysed statistically.Results:①image quality:SSD images of low-dose were better or as well as conventional-dose,and images of high-res-olution were better than standard-resolution.No significant difference in image qualiyy was note with two dose VR images.VR images of high-resolution were better than standard-resolution.②bronchia display:there were no significant difference in SSD and VR images with50and195mAs for displaying bronchi.③lesion detection:there were no significant difference in SSD and VR images of two dose.④dose difference:dose can be decreased greatly with50mAs(13.6-3.6mGy).Conclusion:There is no differ-ence of three-dimensional images of the central airway of low-dose thin MSCT and thoses of conven-tional-dose.Three-demensional images of the central airway with low-dose thin MSCT can be used clinically.
What problem does this paper attempt to address?