Detecting Pulmonary Nodules with Two Reconstruction Slice Thickness in 16-Slice Low Dose Helical CT

史燕杰,张晓锦,李剑颖,于海涛,祝安惠
DOI: https://doi.org/10.13929/j.1003-3289.2010.08.066
2010-01-01
Abstract:Objective To compare the performance of detecting and characterizing pulmonary nodules with 1.25 mm and 2.5 mm reconstruction slice thickness in 16-slice low dose spiral CT.Methods Eighty patients underwent low dose pulmonary CT for health check-up were enrolled because of suspected or diagnosed pulmonary nodules.All nodules were categorized into ≤5 mm(goup 1),5—10 mm(group 2) and ≥10 mm(group 3) nodule.Images with 1.25 mm and 2.50 mm reconstruction slice thickness were generated for each of the patient.The number of nodules,internal structure(existence of calcification and cavity) and peripheral structure(clarities of edge,lobulation and spiculation signs) for each of the reconstruction were recorded.Results There was no statistical difference in the detected number of nodules between 1.25 mm and 2.50 mm reconstruction slice thickness.In Group 1,no statistical difference of detecting spiculation and cavity evaluation existed between the two reconstruction slice thicknesses,but nodule boundary,lobulation or calcification.In group 2 and 3,there was no difference for detecting nodule characterization between the two reconstruction slice thicknesses.Conclusion During low dose CT scanning for detection of pulmonary nodules,2.50 mm can be used for the same image noise as the 1.25 mm reconstruction slice thickness to provide further dose reduction.