Experimental Study on Quality Factor of Superconducting Receiving Coil for MRI

Bai Zhiming,He Yanfa,Wang Yinan,Wang Jinxing
2008-01-01
Rare Metal Materials and Engineering
Abstract:The radio frequency (RF) receiving coil is an important element in magnetic resonance imaging(MRI) system. The quality factor (Q) of superconducting receiving coil was studied experimentally, and the testing principle and method were presented in this paper. A kind of double detecting coil method was used to measure the quality factor of superconducting receiving coil made of Bi2223/Ag tape at room temperature and liquid nitrogen temperature. These testing results were contrasted with those of Cu resonance loop. The results show that the quality factor of superconducting receiving coil is double bigger than that of normal copper coil at room temperature. This indicates that using superconducting material as the RF receiving coil can efficiently enhance the quality factor of resonance loop, and accordingly improve the signal noise ratio (SNR) and image quality of MRI system.
What problem does this paper attempt to address?