Analysis of Target Location Errors in Radiation Treatment of Non-Small-Cell Lung Cancer Using Active Breathing Control System

ZHANG Xian,WANG Jian,ZHU Jiangyi,ZENG Zhaochong,WU Zheng,BAI Chunxue
DOI: https://doi.org/10.3969/j.issn.1008-6358.2012.02.007
2012-01-01
Abstract:Objective:To analyze the comprehensive errors of the active breathing control(ABC) system and the causes when it is applied to three-dimensional conformal radiotherapy(3D-CRT) for non-small-cell lung cancer(NSCLC) and to investigate the feasibility of its application in stereotactic body radiotherapy(SBR) for NSCLC.Methods:A total of 8 patients with phase II-III NSCLC were fixed by the vacuum phantom and treated with 3D-CRT.During the whole process of plan preparation and radiotherapy,ABC system was used to control the relative displacement caused by respiratory movement.Elekta iviewGT system was used to record the displacement of target position in the simulation and treatment.The inter-fractional radiotherapy errors and the intra-fractional radiotherapy errors were analyzed.Results: The overall inter-fractional radiotherapy errors were(1.28±1.02) mm for the X-axis positive direction and(1.07±1.76) mm for the Y-axis positive direction.The overall intra-fractional radiotherapy errors were(1.35±1.63) mm for the X-axis positive direction and(1.38±1.50) mm for the Y-axis positive direction.Conclusions: The inter-fractional radiotherapy errors and intra-fractional radiotherapy errors of 3D-CRT using ABC system were significantly smaller than those of conventional radiotherapy.