Positioning accuracy for lung stereotactic body radiotherapy patients determined by on-treatment cone-beam CT imaging

N D Richmond,K E Pilling,C Peedell,D Shakespeare,C P Walker
DOI: https://doi.org/10.1259/bjr/54341099
Abstract:Stereotactic body radiotherapy for early stage non-small cell lung cancer is an emerging treatment option in the UK. Since relatively few high-dose ablative fractions are delivered to a small target volume, the consequences of a geometric miss are potentially severe. This paper presents the results of treatment delivery set-up data collected using Elekta Synergy (Elekta, Crawley, UK) cone-beam CT imaging for 17 patients immobilised using the Bodyfix system (Medical Intelligence, Schwabmuenchen, Germany). Images were acquired on the linear accelerator at initial patient treatment set-up, following any position correction adjustments, and post-treatment. These were matched to the localisation CT scan using the Elekta XVI software. In total, 71 fractions were analysed for patient set-up errors. The mean vector error at initial set-up was calculated as 5.3 ± 2.7 mm, which was significantly reduced to 1.4 ± 0.7 mm following image guided correction. Post-treatment the corresponding value was 2.1 ± 1.2 mm. The use of the Bodyfix abdominal compression plate on 5 patients to reduce the range of tumour excursion during respiration produced mean longitudinal set-up corrections of -4.4 ± 4.5 mm compared with -0.7 ± 2.6 mm without compression for the remaining 12 patients. The use of abdominal compression led to a greater variation in set-up errors and a shift in the mean value.
What problem does this paper attempt to address?