High-repetition-rate laser ablation for elemental analysis in an inductively coupled plasma with acoustic wave normalization

Ho Ming. Pang,Daniel R. Wiederin,R. S. Houk,Edward S. Yeung
DOI: https://doi.org/10.1021/ac00004a017
IF: 7.4
1991-01-01
Analytical Chemistry
Abstract:A laser ablation system with 100-Hz repetition rate for direct elemental analysis of solids by inductively coupled plasma atomic emission spectrometry (ICP-AES) and inductively coupled plasma mass spectrometry (ICP-MS) is described. Detection limits of 12-mu-g/g and 4 ng/g for ICP-AES and ICP-MS, respectively, are observed. These improvements are attributed to the use of low vaporization laser powers (50 mJ), which may produce a more uniform and finer spray. Precision is improved by a factor of 2 or more by averaging over a larger number of laser pulses within the same measurement and by normalizing the emission signal to the amplitude of the acoustic wave generated during the ablation process.
What problem does this paper attempt to address?