Pump- and Valve-Free Flow Injection Capillary Liquid Electrode Discharge Optical Emission Spectrometry Coupled to a Droplet Array Platform.
Anqin Leng,Yao Lin,Yunfei Tian,Li Wu,Xiaoming Jiang,Xiandeng Hou,Chengbin Zheng
DOI: https://doi.org/10.1021/acs.analchem.6b03185
IF: 7.4
2016-01-01
Analytical Chemistry
Abstract:A. miniature (2.5 cm, length x 2.0 Cm- width x 1.9, cm 'height), low power,(<10 W)-, and capillary liquid electrode microplastna optical emission spectrometer was developed for 'rapid 'determination 'of metallic species in aqueous solutions. The sample Solution can be automatically introdnced, into the source -without 4 pntrip owing to the inherent capillary attraction and the force arising from the solution vaporization induced,by micro plasma. A droplet array was used as, a sampling platfcirtrrto realize flow injection--without using any valve and pump, significantly increasing throughput- to 90 samples h(-1). Sample volume is controlled through the sampling time and reduced to the nanoliter level. With a sampling time of 10 s- (equal to.600 nL), d'etection limits of 30 mu g L-1 (18 pg) and 75 mu g L-1 pg), were obtained for Cd and Hg,-respectively, comparable to those reported- for liquid electrode microplasma- optical emission spectrometry. However-, sample consumption is reduced more than 100-fold, making the proposed 'technique mote suitable for the analysis of elements such as Cd, Hg, Li, Na, and K when sample volumes may be limited. The utility of this system was demonstrated by the determination of Cd and Hg,in blood, real water samples, and Certified Reference Materials (rice powder, GBW07601a, and lobster hepatopancreas, (TORT-3).