Experimental Study of the Spherical Aberrations of Traditional Intraocular Lenses

汤欣,杨瑞波,杨勇,方志良
DOI: https://doi.org/10.3760/cma.j.issn.1008-1801.2007.01.002
2007-01-01
Abstract:Objective To study the spherical aberrations (SA) of traditional intraocular lenses (IOLs) and the effect of SA on visual function.Methods A special optical metric system was set up to measure the wavefront aberrations of IOLs with a Hartmann-Shack aberrometer. All 8 types of traditional IOLs were divided into 2 groups: group 1 consisted of 4 types of IOLs made of different lens materials; group 2 consisted of 4 types of IOLs with different designs. IOL aberrations were measured continuously under 3 mm, 4 mm, 5 mm, 6 mm, and 7 mm apertures. The data was then statistically analyzed.Results All 8 types of traditional IOLs had positive SAs. With increases in aperture size, the SAs of the two groups increased accordingly. There were positive correlations between aperture size and SA (r=0.6465,0.7872;P0.001). SA made up more than 50% of the total aberrations (TA) with 3~7 mm apertures in both groups. The average TA increased gradually with 3~6 mm apertures. With 3~7 mm apertures, the number of IOLs that matched Rayleigh criterion were 8, 5, 3, 2, 2, respectively.Conclusion Spherical IOLs produced positive SAs, so they are unable to compensate for positive corneal SAs, which are similar to the aging crystalline lens. The SA of the IOL was the main part of the total aberration in the axial imaging system. The quality of the IOL’s optics was perfect with a small aperture, especially smaller than 4 mm. As the size of the aperture increased, the quality of the IOL’s optics decreased.
What problem does this paper attempt to address?