Precision of DXA for the Measurements of Human Body Composition

XIAO Ze-yu,GUO Bin,GONG Jiang,TANG Yong-jin,XU Hao
DOI: https://doi.org/10.3969/j.issn.1000-9965.2012.06.009
2012-01-01
Abstract:Aim:To measure the precision of body composition measurements and fat distribution in human using dual energy X-ray absorptiometry(DXA),and evaluate its reliability. Methods: Ten consecutive total body scans were conducted with DXA(GE Lunar Prodigy)in 20 healthy adult volunteers(10 males,10 females).The software(enCORE,version10.50.086)was used to analyze total bone mineral content(TBMC),total lean mass(TLM),total fat mass(TFM),total body fat percentage(%TBF),android fat percentage(%AF),gynoid fat percentage(%GF),the precision was represented as the coefficient of variation of root mean square(RMS-CV),standard deviation of root mean square(RMS-SD). Results: RMS-CV were 1.05%,1.95%,0.93%,1.98% and RMS-SD were 0.025,0.234,0.377,0.353 for TBMC,TFM,TLM and %TBF.RMS-CV were 2.51,1.80 and RMS-SD were 1.153,1.101 for %AF and %GF. Conclusion:DXA provided excellent precision for the measurements of body composition and fat distribution.