The Study of the Precision and Accuracy of Quality Control in DXA Bone Mineral Densitometry

GONG Jian,XU Hao
DOI: https://doi.org/10.3760/cma.j.issn.2095-2848.2005.06.003
2005-01-01
Abstract:Objective To study the precision and accuracy of quality control(QC)in dual-energy X-ray absurptiometry(DXA)bone mineral densitometry so as to raise the reliability and necessity of the re- sults.Methods ①Short-term precision trial:30 people and 30 SD male rats were chosen,and a precision trail was performed.Each people was scanned twice and reposited in next study.The precision and the least significant change(LSC)of each examinated region were calculated.The short-term precision trail of the rats was performed in the similar way.②Accuracy trial:measured the body phantom supplied by factory daily, and compared the results with real value,then calculated the accuracy and correction factor.A Shewhart chart was set up based on average values.Results ①People's coefficient of variation(CV)and LSC in the lumbar and proximal femur were 0.7%~2.2% and 0.018~0.048 g/cm~2.Rats' whole body short-term precision was 0.9%.②The average accuracy of DXA densitometer was-0.81%,the conrrection factor was 0.992.The average bone mineral denisity measured in successive 25 d was 1.244 g/cm~2,the standard de- viation(SD)was 0.008.Conclusion The precision and accuracy trail can help to get the information a- bout the working state of the instrument and to analyze the measured results,and can effectively raise the re- liability of the measure.