Calibration of Neutron Detection Efficiency of Li-glass Detector

ZHANG Qiwei,HE Guozhu,RUAN Xichao,LI Xia,REN Jie,ZHOU Zuying
DOI: https://doi.org/10.11804/nuclphysrev.30.02.156
2013-01-01
Abstract:In order to accurately measure the neutron capture cross section in the energy range of keV~MeV,4π gamma-ray total absorption facility(GTAF) is being constructed at China Institute of Atomic Energy(CIAE).The lithium glass detector will be used as a neutron beam monitor for GTAF.The detection efficiency of the lithium glass detector at two incident neutron energy points(250,565 keV) was calibrated in 5SDH-2 accelerator,and the relative detection efficiency was simulated by GEANT4 and MCNP code.By the normalization of the experimental data and simulation result,the neutron detection efficiency curve of the lithium glass detector between 10 keV and 1 MeV was obtained.This work will be important to convert the Time-of-flight spectrum that be measured by Li-glass detector to the energy spectrum of neutron beam,and provide the new method for calibration of detection efficiency.
What problem does this paper attempt to address?