Application of 4D-CT in CyberKnife for Lung Tumor

WANG Tao,SHEN Jun-shu,ZHU Xi-xu,LI Bing
DOI: https://doi.org/10.7687/j.issn.1003-8868.2013.08.067
2013-01-01
Abstract:Objective To determine the motion vectors of lung tumors in three-dimensional direction with 4D-CT to provide CyberKnife treatment reference for PTV margin.Methods Eleven cases of lung cancer patients were scanned with PHILIPS Brilliance Big Bore 16-slice spiral CT in Pulmo RETRO 4D mode with 10-phase images.The motion vectors and sizes of lung tumors were measured in Tumor LOC movie mode.Results A total of 5 cases with movement less than 5mm in threedimensional direction,including 3 cases with upper lung cancer,1 case of right-lower lung cancer and 1 case of right middle lung cancer,were treated with CyberKnife using X-Sight Spine tracking.Conclusion By using 4D-CT to determine the scope of the tumor three-dimensional movement and enlarging the GTV to PTV,some patients with lung tumors can be treated with CyberKnife by X-Sight Spine tracking,who are incompatible to gold seeds implantation or X-Sight Lung tracking.