Measurement System Design of Electrophoresis Microchip Conductivity Detection Based on SoPC Technology

LIAO Hong-hua,WANG Jun,ZHOU Wen-li,CHEN Jian-jun,YI Jin-qiao,YU Jun
DOI: https://doi.org/10.3969/j.issn.1002-1841.2008.10.025
2008-01-01
Abstract:A kind of measurement system design of capillary electrophoresis microchip based on SoPC technology was described.The synchronous orthogonal signals were obtained by using MAX038 high frequency waveform generator,the impedance was converted to voltage based on the dua1 differential subtraction,the magnitude and phase of the signal conductivity were detected by using the dual lock-in amplifiers,the embedded software core processor.NiosII,acted as the key and kernel controlling module in FPGA,the FFT controller and the high speed signal sampling controller,which were defined by SoPC Builder software,and the ultra-high speed,real-time FFT transform module,which was designed using FFT V2.1.3 IP core,and the time domain and frequency-domain analysis of capacitive coupled contactless conductivity in electrophoresis microchip were achieved.The architecture of software and hardware in measurement system and the custom FFT instruction for NioSII embedded processor were discussed.The simulation experimental results show that this project scheme can process the complex data computing,having quick and real-time processing ability,it can effectively ensure system anti-jamming ability and the accurate measurement.
What problem does this paper attempt to address?