Determination of b value in diffusion weighted imaging of orbital masses

蒯新平,王胜裕,陶晓峰,刘士远,胡蓓西,丁庆国
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2013.10.004
2013-01-01
Abstract:Objective:To study the influences of different b values on the image quality of DWI and ADC value of orbital mass.Methods:Routine MRI and DWI were performed in 81subjects with intra-orbital mass before surgery with the b value=0and/or 400,700,1000s/mm2.SE-EPI sequence with fixed parameters combination was used for DWI.The signal noise ratio(SNR)of lesion,contrast-to-noise ratio(CNR)of lesion/white matter of contra-lateral temporal lobe,ADC value of lesion,ADC ratio(ADCR)of lesion/white matter of contra-lateral temporal lobe were calculated with different b values(400,700,1000s/mm2).Results:The SNR of intra-orbital tumor on DWI(b=400s/mm2,700s/mm2,1000s/mm2)was85.10±48.90,61.46±37.00,42.92±25.53respectively(χ2=125.654,P=0.000),with significant statistic differences.The CNR between the mass and white matter of contra-lateral temporal lobe with different b values was 28.36±26.67,20.11±17.19,16.59±15.29respectively(χ2=14.519,P=0.000),with significant statistic differences.The ADC values were(1.42±0.44)×10-3 mm2/s,(1.25±0.43)×10-3 mm2/s and(1.16±0.40)×10-3 mm2/s with b=400s/mm2,700s/mm2and 1000s/mm2,respectively(F=7.932,P=0.000),with significant statistic differences.The ADCR was(1.47±0.45),(1.39±0.46)and(1.37±0.46)with b=400s/mm2,700s/mm2 and 1000s/mm2,respectively(F=1.199,P=0.303),with no statistic difference.Conclusion:The optimal b value in demonstrating orbital mass on DWI was 700s/mm2.
What problem does this paper attempt to address?