Abstract:Purpose:To study the effect of different b values on signal intensity(SI),ADC and the quality of image.Methods:Three different weightings(b=1000,2000,3000 s/mm2)were applied in thirty normal brains in diffusion weighted imaging on 1.5T MR machine.The averages of SI and background noise were measured.The averages of SNR and CNR were computed from these data and compared.Results:The measured SI,ADC and SNR were significantly decreased accompanied with increase of b value(P0.05),but the CNR was significantly increased at posterior limb of internal capsule and decreased in gray matter ROIs.Conclusion:The strength of b value influences measurement of SI,ADC and the quality of image.Different b value imaging appears different brain structure information.