Relationship between CBCT Image Quality and Image Dose of Accelerator's Cone Beam CT Equipment with Bowtie-filter

PENG Guang,XU Qing-feng,CHEN Gang,BAI Sen
2010-01-01
Abstract:Objective To explore the influence of bowtie filtration on absorb dose and half-value layer inaluminum (Al) of cone-beam CT, estimate the image dose under different scan protocol, and establish the relationship between the image quality and the scan protocol after using F1. Methods Piranha was used to measure the HVL. Dose measurements were performed with an 0.6 cc Farmer type ionization chamber with a 30 cm-diam cylindrical shaped water phantoms in 100 and 120 kV with a series of mAs and FOV. CNR, noise and uniformity were measured on the Catphan503 images. Results HVL increased 0.77-0.92 mmAl where XVI generally had more penetrating beams at the similar kV settings. Scanning dose significantly reduced, the center point decreased 22%-29%, the edge with a decrease of 41%-45% which meant a very significant reduction in skin dose. Image quality improved with mAs increase. The spatial resolution mainly changed with FOV. But generally can identify 1-2 mm-diam objects, fully meet the clinical requirements of identify small bone structure and marker. Through this clinical investigation, low-dose CBCT images in chest, such as 100 kV, M20, and 0.5 mAs scanning protocol appeared to be an optimal settings. Abdomen image needed a higher dose to reach the requirements of registration. Conclusions Using F1 under the premise of improving the image quality then reducing the scanning dose and using a new scanning sequence can balance the image quality and scanning dose.