Study on the impact of bowtie filter on photon-counting CT imaging

Xin Zhang,Jixiong Xie,Ting Su,Jiongtao Zhu,Dongmei Xia,Hai-Rong Zheng,Dong Liang,Yongshuai Ge
DOI: https://doi.org/10.1088/1361-6560/ad8858
IF: 3.5
2024-10-18
Physics in Medicine and Biology
Abstract:Objective: The aim of this study was to investigate the impact of the bowtie filter on the image quality of the photon-counting detector (PCD) based CT imaging. Approach: Numerical simulations were conducted to investigate the impact of bowtie filters on image uniformity using two water phantoms, with tube potentials ranging from 60 to 140 kVp with a step of 5 kVp. Subsequently, benchtop PCD-CT imaging experiments were performed to verify the observations from the numerical simulations. Additionally, various correction methods were validated through these experiments. Main results: It was found that the use of a bowtie filter significantly alters the uniformity of PCD-CT images, depending on the size of the object and the X-ray spectrum. Two notable effects were observed: the capping effect and the flattening effect. Furthermore, it was demonstrated that the conventional beam hardening correction method could effectively mitigate such non-uniformity in PCD-CT images, provided that dedicated calibration parameters were used. Significance: It was demonstrated that the incorporation of a bowtie filter results in varied image artifacts in PCD-CT imaging under different conditions. Certain image correction methods can effectively mitigate and reduce these artifacts, thereby enhancing the overall quality of PCD-CT images.