The Study on Dose Measurement for CBCT of XVI System

查元梓,张松方,沈奕晨,吴国华
DOI: https://doi.org/10.3969/j.issn.1005-202x.2010.05.006
2010-01-01
Abstract:Objective:To explore a dose measurement method for CBCT scanning of XVI system.Methods:In the study,we measured clinical used protocols in our department(head and and protocol),uniform-density cylindri-cal acrylic phantom was used for all measurements.The phantom included ion chamber placement holes in the center and at periphery,absolute dose of the total 5 points can be separately achieved,computed tomography dose index(CTDIw) was used to evaluate average dose from the scan.Results:(1)The average dose is closely related to mAs;(2)The average dose ranges from 0.1 cGy to 5.39 cGy with the highest dose measured using the protocol with the phantom;(3)For different kinds of fil-ter,F0 and F1,dose difference is obvious.Conclusions:To reduce extra dose to patients,for pediatric as well as and patients,it's advised to adopt head and neck protocol and F1 filter.For chest and pelvis scan,it's also advised to use F1 filter.