The Fast and Sensitive System for Determination of Blood Lead Based on the Integrated Film Electrodes Without Removal of Oxygen

Wei Li,Jilie Kong
DOI: https://doi.org/10.1080/00032710701566792
2007-01-01
Analytical Letters
Abstract:A novel fast system based on the integrated film electrodes was created. The resulting system was successfully applied in the detection of blood lead, by means of differential pulse anodic stripping voltammetry in tiny amount of volume, similar to 50 mu 1, code-posited with Hg without removal of oxygen. Under the optimum conditions, the peak current was linear to Pb2+ concentration from 1.2 to 70 mu g/d1, and detection limit was 1.2 mu g/d1. Blood lead was detected with calibration curve, y = 9.726E-8 + 9.959E-8 x (R = 0.998). Compared with other methods reported, this method was more rapid, low-cost, and could meet the requirement of real-time analysis.