The fast system for analyzing blood lead based on the integrated film electrodes

Li Wei,Shi Mian-hong,Kong Ji-lie
DOI: https://doi.org/10.3969/j.issn.1008-2298.2006.04.003
2006-01-01
Chemical Sensors
Abstract:At present,most domestic methods for analyzing lead concentration of blood suffer from some disadvantages,such as large acquirement of sample amount,high price of apparatus,as well as reiteration of digesting samples,multi-steps and time-costing process,which greatly increase the cost of the analysis.Employing differential pulse anodic stripping voltammetric (DPASV) and integrated three-electrode system,which combines the technology of sensor,modern printing and micro-machining,we can actualize direct microanalysis of lead in one drop of blood (about 50 μ L) without digestion in 3 min.The linearity of the standard working curve is good in the range of 1.2~ 70 μ g/dL at pH1.1.The correlation coefficient is 0.998.Such rapid,accurate,low-costing process can meet the requirement of real-time analysis.