Comparison of the Results of Potassium & Sodium Determining in Mixed Serum by Different Electrochemistry Measurements

HU Jin-cao,GUO Qun,GU Guang-yu,HUANG Ai-jun,ZHANG Kui
DOI: https://doi.org/10.3969/j.issn.1671-7414.2008.02.015
2008-01-01
Abstract:Objective Comparison of the results of potassium & sodium determining in mixed serum by direct potentiometric method,indirect potentiometric method and dry chemistry direct potentiometric method.To explore the importance of calibration and comparison in electrochemistry analysis.Methods Calibrated Vitros 250,AVL 9180,EX-Z & Hitachi 7600-020 biochemistry analysators.Then evaluated the precision,the recovery,the linearity and the contrast test.Finally,the results were analyzed with statistics.Results The results of the analysators showed the good comparability and correlation,statistical significance was less different obviously.Conclusion For showing good accuracy and consistency,the calibration and contrast test for analysators must be done well in routine work.