Evaluation of Sysmex UF-500i Automated Urinary Sediment Analyzer in Screening Urinary Tract Infection

WANG Jin-hua,ZHANG Lei,LI Wei,WANG Xiang-ling,FENG Nan,ZHU Jian-hong
DOI: https://doi.org/10.3969/j.issn.1672-9455.2012.03.017
2012-01-01
Abstract:Objective To study the value of UF-500i urinary sediment analyzer on assisted diagnosis of urinary tract infection.Methods 345 specimens of midstream urine were examined for leukocyte and bacteria counts by UF-500i urinary sediment analyzer.However,bacterial culture was conducted.According to the results of bacterial culture,ROC curve was drawn with SPSS13.0 software,the cut-off value of leukocyte and bacteria counts for diagnosis of urinary tract infection were found out,and its sensitivity,specificity,positive/negative predictive value,false positive/false negative rate and accuracy were calculated.Results The positive rate of urine culture was 27.8%,and the most common pathogen was Escherichia coli.The obtained cutoff values were 130.2 bacteria/μL and 29.8 leukocytes/μL.Conclusion The UF-500i automated urinary sediment analyzer can be adopted to conduct simple,rapid and reliable urine screening tests,its bacterial and leukocyte counts may be used as the better indexes for diagnosing urinary infection.