Adding Auxiliary Electrode—An Effective Method for Enhancing Signal‐to‐Noise Ratio in Nanospray Mass Spectrometry

XJ Xu,JJ Zhai,WQ Shui,GB Xu,PY Yang
DOI: https://doi.org/10.1081/al-200031971
2004-01-01
Analytical Letters
Abstract:A simple arrangement for improving signal-to-noise ratio (S/N) in a nanoelectrospray procedure using an auxiliary electrode has been investigated. The auxiliary electrode (aluminium plate in the present experiments) with different size and shape was placed near the home-made nanoelectrospray tip and various electrical potential was added, which could dramatically disturb nanoelectrospray properties. This arrangement was employed to enhance S/N of tips spraying reserpine solution at different concentrations under the control experimental conditions. The comparison of S/N of nanoelectrospray with and without auxiliary electrode gives an explicit trend that auxiliary electrode could improve sensitivity of nanoelectrospray to at least twofold.