Calibration method for planar SQUID gradiometers based on the magnetic gradient tensor components
Jun Wu,Liangliang Rong,Haozhen Jia,Guofeng Zhang,Longqing Qiu,Hui Dong,Yifeng Pei,Shulin Zhang,Xiaobing Yuan,Baoqing Li
DOI: https://doi.org/10.1088/1361-6501/acdb8b
IF: 2.398
2023-06-06
Measurement Science and Technology
Abstract:The accuracy of Tesla/Volt calibration factors for SQUID gradiometers is essential to take advantage of magnetic gradient tensor (MGT), but the accuracy of the published calibration procedures rang from tenths of per cent to a few per cent. In this paper, we propose an efficient calibration method for planar SQUID gradiometers based on the MGT components. The factors affecting the calibration accuracy and the optimal inclination angle for calibration were investigated by analyzing the response characteristics of planar SQUID gradiometers to the MGT components. By superposition of square coils and gridding of pickup loops, numerical models of the response of planar gradiometer in different types of Maxwell coils were built to simulate theoretical calibration accuracy and calibration errors due to position, attitude and imbalance. We built a calibration setup to achieve high precision simultaneous calibration of multi-channel planar SQUID gradiometers and investigated the effect of eddy current on the calibration accuracy. The results show that simulation and experimental results were in general agreement by taking into account various error factors, and a calibration error of about 0.4% was achieved.
engineering, multidisciplinary,instruments & instrumentation