Optimal Parameters of MR Diffusion Weighted Imaging with Background Body Signal Suppression on Pulmonary Diseases

Wu Huawei,Lu Qing,Cheng Jiejun
DOI: https://doi.org/10.3969/j.issn.1006-5741.2007.01.006
2007-01-01
Abstract:Purpose:To study the optimal technique of MR diffusion weighted imaging with background body signal suppression(DWIBS) on pulmonary diseases according to the comparison of image quality with different parameters.Materials and Methods:Twenty patients with pulmonary diseases underwent DWIBS on a 3.0T MRI, using different parameters including breath-hold vs. free breathing methods, different b values(0, 250,500,800),SE-EPI SPIR vs.STIR-EPI,different thickness(4mm vs,9mm). ADC value, quality index, SNR and AppCNR were analyzed for different parameters.And the feasibility of high resolution MPR and MIP images were evaluated. Results:The quality index, SNR and AppCNR of free breathing methods were better than breath-hold methods(P<0.01). The quality index and AppCNR of b500 was the highest among different b value groups(P=0.000).The quality index and AppCNR of STIR-EPI were better than SPIR(P<0.05).The quality index of 4mm slice thickness free breathing scans was better than 9mm slice thickness free breathing scans(P<0.05), but SNR and AppCNR had no significant difference between two scan methods(P>0.05). The distinction of quality index ,SNR and ADCs value were not significant among different TR groups(P>0.05). The AppCNR of TR 6000 and TR8000 was higher than TR4000 group(P<0.05).Conclusion: High quality DWIBS can be achieved on 3.0T MRI with optimal parameters( free breathing scans, b500, STIR-EPI and TR 6000) .
What problem does this paper attempt to address?